Description
Use this pop-up to view and edit order choice information for a particular instrument test from the Instrument Types page.
Access
AdminCenter > Instrument Setup > Instrument Types > Test Setup
Technical Manual
Tests Pop-up
Up and Down Arrows | Select the arrow buttons to reorder the tests on the pop-up. |
---|---|
Order Choice | Select inside or enter text in the search field to select an order choice for the test. Select the magnifying glass icon to manually search for an order choice. Select X to remove text from the field. |
Set ID | Select inside or enter text in the search field to select a set ID. This field allows an identifier to be specified for the test set. This setting is not required to be populated for the test set. If an unsolicited test results message is received from an ODE with a value specified in OBR-4, the system will look for the matching test set to use by seeing if the OBR-4 value matches the Set ID for any of the test sets on the instrument type. If OBR-4 is blank, or does not match to any existing test sets, the system will continue with its previous functionality of selecting the test set based on the tests present in the message. If a solicited test results message is received from an ODE with a value specified in OBR-4, the system will look for the matching test set to use in the same way as unsolicited. If a matching test set is found for the value, and an order choice on the solicited order can use that test set, the matching test set will be used. If a matching test set is found for the value, but no order choice on the solicited order can use the test set, the results will not parse to the solicited order. If no matching test set is found for the value, the system will continue its previous functionality of selecting the test set based upon the tests present in the message. |
Add New Test | Select to add a new test on the pop-up. This button will also be present underneath the individual test sets for easy access. |
Add Existing Test | Select to add an already built test on the pop-up. This button will also be present underneath the individual test sets for easy access. |
Delete Order Choice | Select to remove the order choice from the pop-up. This button will also be present underneath the individual test sets for easy access. |
Collapse Tests | Select the link to collapse the individual tests to create more space on the pop-up. This button will also be present underneath the individual test sets for easy access. |
Hide Tests | Select the link to completely hide the individual tests from the page. This button will also be present underneath the individual test sets for easy access. |
Add Order Choice | Select to add a new order choice to the instrument. |
Delete Test | Select to delete the test from the pop-up. |
Non-Reportable | Select the checkbox to specify that the test is a non-reportable test. |
Test is mandatory | Select the checkbox to have the test be considered mandatory for order choices when the Can be approved with setting is set to All mandatory results present. Note that the Test is mandatory checkbox must be set separately for each instrument type/order choice combination that a test appears on. |
Add Reference Range | Select to add reference range information to the order choice test.
|
Override with instrument reference ranges | Select the checkbox to override the order choice with the instrument reference ranges. |
Name | Enter the name of the test. This is a required field. |
Abbreviation | Enter an abbreviation for the test. |
Precision | Enter a number for precision for the test. |
Identifier | Enter an identifier for the test. |
LOINC | Enter a LOINC code for the test. |
Units | Enter the units for the test. |
Enterability | Select one of the following options from the drop-down list to specify how text can be entered:
|
UCUM Units | Enter UCUM units for the test. |
Host Codes | Select to display the Test Host Codes pop-up. The pop-up table displays the host name and the host code.
|
Enterability | Select an option from the drop-down list to specify how information can be entered:
|
Choice List | Select to display the Choice Result List pop-up to add a predefined result list.
|
UCUM Units | Enter text in or select in the search-as-you-type field to select UCUM units for the order choice. |
Formula Settings | Select to display the Formula Settings pop-up.
|
Qualitative Normal Values | Select to display the Qualitative Normal Values pop-up to add a qualitative normal value.
|
Qualitative Critical Values | Select to display the Qualitative Critical Values pop-up to add a qualitative critical value.
|
Qualitative Abnormal Values | Select to display the Qualitative Abnormal Values pop-up to add a qualitative abnormal value.
|
Delta Settings | Select to display the Delta Settings pop-up to configure delta settings.
|
Westgard Rules | Select to display the Westgard Rules pop-up to define Westgard Rules evaluations for QC results. The following rules are available: 1 2S , 1 2.5S , 1 3S , 1 3.5S , 2 2S , 4 1S , R 4S , 10 1S , 10 X , and 7 T . Select the appropriate checkbox for the rule(s). Select Save to add the Westgard Rules to the Tests pop-up. Select Cancel to close the pop-up without adding Westgard Rules. The button will have green, italicized text when Westgard Rules have been added. Westgard Rules will be evaluated when a numeric result is received/entered and when approving results. Results that have violated any Westgard Rules will have a W flag in the Flags column on the Review Quality Control pop-up on the Review QC page. Hovering over the flag will display why the violation occurred. Any Westgard Rules violations that were triggered by the approved result will be displayed on the Review Quality Control pop-up's new Violations field via Status. See the Review QC page for more information. |
Save | Select to save the changes made on the pop-up. |
Cancel | Select to return to the Instrument Type details page without saving changes. |